FAQ Resources
Filter by Product/Type
Or Filter By Language
How long do we have to put the dental models in Asiga flash to cure?
One 180 sec cycle should work for WMC model resins. If the surface still feels a little tacky you can run a second 180 sec cycle.
How often do we need to clean the vat and maintain to zero position?
Only after a failed print (tray cleaning process) or a “Failed to approach – Target not reached in XX seconds” (debris in tray or calibration joint has shifted)
Help! I had delamination while printing on an ASIGA Pro2. How can I prevent that from happening?
If you experience delamination while printing on the Pro2 printer, recalibrate the scanner, clean the glass slider and glass on the resin catch tray. Then check for debris on the projector lens and clean. If the problem persists, check the glass slider itself. We have seen a few cases where the glass on the slider breaks loose from the slider itself so the glass plate is no longer adhered to the slider. If you experiences this, follow the steps below to fix the issue.
1. Remove glass slider from printer.
2. Make sure the frosted side of the glass is facing upwards.
3. Open a warranty claim with Asiga if unit is still under warranty. This is done on the Asiga website, www.Asiga.com.
4. Take glass and slider to a windshield repair shop to have the glass re-adhered to the slider. Make sure the material used to adhere the glass to the slider is not higher than the top of the glass, this can cause it to rip a resin tray.
5. Re-install the slider into the printer. Inspect the glass slider to make sure it was re-adhered correctly. Follow instructions to reinstall the glass slider or follow removal instructions in a backwards manner.
6. Run a test print to see if the slider is fixed and resolves the delamination issues.
I had a failed print, now what?
We have had several calls recently where labs have had failed prints. They are either trying to pick the cured resin off the bottom of the resin tray when calling or attempting to start another print immediately after a failed print without removing the cured resin from the bottom of the resin tray. Doing this causes “approach errors”, due to an obstruction being detected. To remedy this issue, do the following steps when you get a ‘failed print’.
1. Remove any cured resin from build platform, and clean thoroughly with isopropyl alcohol.
2. Run Clean Tray Function
3. Check bottom side of resin tray and glass for dirt/smudges/residue. Wipe any resin residue or smudges from bottom side of build tray and glass of the printer.
4. Re-insert resin tray, and build head.
5. If failed prints continue, call us at Whip Mix Tech Support.
How much Resin Should I Put into the Tray of the Printer?
For a MAX user 0.5Kg is Ok. We can use 200-300 gm (0.20- 0.3 kg) of material to print a job because the MAX has a small tray. (Print Trays come in Different sizes and Number of Uses)
For a PRO user, 0.5 kg is not enough because of the size of the tray. The pro tray needs 700-800 gm of material to do a print a job.
There are icicles or “flash” around the edge of the build plate on an Asiga printer. What can I do?
If you are noticing “flash” or “icicles” around the edges of the build plate on your printer; this is most often caused by the projector nearing the end of its life expectancy. Once the projector starts to go bad, projected light will begin to scatter on the edges of the image. Contact Whip Mix Technical Support for further assistance.
How do I update Asiga Composer Software?
1. Log into Asiga.com account
2. Select on my products and click on software
3. Select and download the version of software that you desire
4. Locate the downloaded installer and run the installer as administrator
5. Select next through the prompts and enjoy the update
How do I update the Asiga Printer Firmware?
1. Log into Asiga.com account
2. Select on my products and click on firmware
3. Select on your printer
4. Select on the version of the firmware you wish to download
5. Open composer software
6. Open printer web interface (the world icon)
7. Select the update firmware tab
8. Select ‘choose file’ and locate the firmware you downloaded (look in ‘downloads’ folder) select the firmware and then select ‘OK’
9. Select upload- IMPORTANT stay on this screen until upload states successful
10. Go to printer and select on the systems tab on the touch screen
11. Select firmware
12. Select update firmware
13. The printer will reboot
14. Enjoy the update
How do I Verify my Calibration on the Asiga MAX?
Calibration is important to the success of a print build. Following this guide will let you check your Asiga MAX to see if the calibration of the build plate has shifted. This could be one cause for the error, “Failed to reach target in 60s.”
1. Remove the resin tray and clean the build head.
2. Put the build head in the printer and tighten the black build head knob.
3. Zero the position encoders. Go to maintenance > Position Encoders > Zero Position Encoders > Press the “Zero” button. The numbers on the left of the screen should be +- 0.010. It is ok if they fluctuate as long as the fluctuation is small.
4. Press and hold the down arrow until the build head stops moving. Now check the values on the left. If you see any large values, the Zero Position calibration was done incorrectly and needs to be done again. If the values stay small then the calibration of the Zero Position is good.
What is the best Asiga Printers light source?
Asiga Printers are available with 405nm (nano-meters) or with 385nm. The 405nm machines are multipurpose. The 385nm machines have the UV light availability, which is necessary for curing clear materials.
Build plate size is important. What size build plate is available on the printers?
Asiga printers have build plates that vary depending upon which printer you use. There are five available build plate sizes:
71mm x 40mm (Pico2 (HD)
96mm x 54mm (PRO 2 (50)
115mm x 64mm (Pro2 (62)
119mm x 67mm (Max)
144mm x 81mm (Pro2 (75)
Which Resins are available to use with the Asiga printer?
Asiga printers are “Open Source” printers and can use any of the light cured resins. You may need to adjust some of the print parameters and you may see a difference in accuracy depending upon the manufacturer of the resin material.
Whip Mix does not have all of the parameter numbers for every print resin on the market and would not. However checking with our Digital Technical Support department before you print may help with your ultimate success.
Resin materials will change the more we use them in our industry. Expect to change some of your numbers over time.
Can we leave VeriModel liquid in the vat for the next build?
Yes, but stir it before printing if it has been sitting for an extended period of time. If you notice that there is sediment on the bottom of the tray then it needs to be mixed.
How much VeriModel should we put in the vat? We cannot find the maximum we can put.
1/3rd to half. There is not a mark on the tray.
Is it possible to re-use the alcohol after we washed models?
To some extent, if the models still have resin residue on them after the clean alcohol bath then you need to change the alcohol. At this time the alcohol can’t be filtered and reused when it becomes too dirty to effectively clean printed parts.
What kind of alcohol grade (ex: 70%UPS) should we use to wash the dental models? Is it possible to dilute the alcohol?
Post processing instructions and recommendations are material specific and should be supplied by the resin manufacturer. For WMC resins, we recommend the highest percentage you can get above 90%.
How do I find Historical Scanner Calibration data?
By following the path below, users can find the historical calibration data. This information is used to track employee calibration schedules for the scanner.
C:\Users\[CURRENT WINDOWS USER]\AppData\Roaming\3Shape\ScanServer
How do I use the scan setup for 3Shape lab scanners in ‘Plab’?
TO OPEN PLAB
1. Shut down ScanSuite;
2. Open Plab from C:\Program Files\3Shape\ScanSuite or C:\Program Files (x86)\3Shape\ScanSuite depending on the installed software version
3. Use password: unique
THREE TYPES OF SCAN SETUP
1. Auto setup – performed by inserting the dot plate and pressing “Auto setup”. For D2000 scanners, auto setup is performed for left and right axis separately.
! NOTE: Auto setup is available only for the scanners with the single calibration object.
2. Open setup – manual axis adjustment; keeps the current definition file, updates only the axes position values.
3. New setup – manual axis adjustment; resets the current scanner’s definition file, including color calibration values for the color texture scanners.
! NOTE: New setup can’t be done for E scanner models.
! NOTE: Scanner must be calibrated after any scan setup.
See also: How to restore color calibration after new setup.
PROCEDURE OF MANUAL AXIS ADJUSTMENT
Once entered to the axis adjustment window you should change the position of each axis by pressing arrow buttons until the axis is set to a correct position:
< and > buttons will sightly move the axis in the chosen direction;
<< and >> buttons will move the axis farther in the chosen direction.
CORRECT POSITIONS OF SCANNER’S AXES
1. Linear axis (single line scanner) – the light line should be in the middle of the rotation axis.
2. Linear axis (multiline scanner) – the line # 14 should be in the middle of the rotation axis; approximately four (4) lines should be on both sides of the rotation axis. For D2000 linear axis must be aligned with left rotation.
3. Rotation axis (on D scanners) – flat side of the rotation axis must be parallel to the edge of the swing.
4. Rotation axis (on E scanners) – white line on the height adapter must be in the front; center blue line should coincide with the horizontal line on the scan plate.
5. Swing axis – must be horizontal (parallel to the table surface). On E scanners horizontal position can be checked with a level tool.
How often should I calibrate my Scanner?
Calibration is usually recommended to be done at least once per week, usually on Mondays is the easiest for end users to remember. However depending on the operating environment it could be required to calibrate more than once per week if the scanner temperature is detected to be more than +/- 5 Degrees from the previous calibration temperature.
What is needed for using scans from a 3M True Definition scanner with your 3Shape system?
You will need to make sure that 3M shows you how to export an un-sectioned scan in 3oxz format from their 3M Connection Center software. Once you have those files you would need to have either the Model Builder Trios and Third Party Scans module or the Model Builder STL module. You also have the option of using CAD points instead of buying the model. This is a pay as you go method so you will not have the upfront cost of buying a module. The downside to CAD points is that you have to monitor how many you have so that you do not start a case without the necessary CAD points to finish it. With CAD points the auto-save feature in 3Shape is disabled so you have to finish the case before closing it or you will lose the design. If you decide to get one of the modules, I would recommend staying with the third party one since it will be cheaper upfront. You could always upgrade to Model Builder STL later if you needed it and the cost is be prorated.
What is the best way to clean a 3Shape Scanner?
3Shape has directly addressed this issue in their information as follows, “customer may wipe out any dust or debris from the inside of the scanner with a damp cloth, making sure to never spray any cleaners or solvents directly into the scanner. The customer should never attempt to clean the camera lenses internal to the scanner.”
Please be sure to follow this information and do not deviate. If you have questions please contact Technical Support.
3Shape will not open. An on the screen states: “Problems occurred while reading Dental System settings file: Could not read archive file C:\3Shape Configuration\DentalSystem\Dental System Control Panel\DentalSystemSettings.3ml. Error msg::Unhandled Archive Type”.
1. Backup Current .3ml file
Navigate to the folder location in the error message and change the name of the existing DentalSystemSettings.3ml to DentalSystemSettings-Backup.3ml.
2. Run the installer.
Run the 3Shape Dental System Installer that matches the version they currently have installed. This will regenerate a new DentalSystemSettings.3ml file.
Can I use the same 3Shape Communicate account on more than one Dental System installation?
There should be one 3Shape Communicate account used per one Dental System standalone installation.
Having same 3Shape Communicate account configured (also same 3Shape Communicate Inbox folder) on several installations is not supported and may cause various problems, like the following,
Order duplicates arriving into 3Shape Communicate Inbox;
1. “The order is not up to date. Please get updates from the server and try again” error message when trying to accept the orders in 3Shape Communicate Inbox.
2. “Unsupported material file version: 20xx-1” error message when having the same 3Shape Communicate account configured and the same 3Shape Communicate Inbox folder configured for several PCs with different DS versions.
The supported configuration is to have several PCs using the same 3Shape Communicate account only as a server-client installation.
Help! 3Shape Error Message: “ScanIt Dental communication error. Error connecting to ScanIt Dental”
When trying to start scanning in ScanIt Dental, there is an error message: “ScanIt Dental communication error. Error connecting to ScanIt Dental”.
Possible solutions:
Solution 1
1. Close all 3Shape applications;
2. Rename folder C:\ProgramData\3Shape\ThreeShape.ScanItDental
3. Rename ‘user.config’ file in the relevant Dental Desktop Version folder
4. Restart Dental Manager.
Solution 2
1. Rename folder C:\Program Files\3Shape\ScanItDental
2. Rename folder C:\ProgramData\3Shape\ThreeShape.ScanItDental
3. Run the installer of the same Dental System version on top of the existing installation.
If it doesn’t help, try:
1. Installing available Windows Updates;
2. (Re)installing SlimDX Runtime .NET 4.0 (get installer here);
3. reinstalling .NET framework;
4. temporarily disabling antivirus;
5. making clean installation of Dental System.
Help! My screen says 3Shape “Cannot Find Network Folder Location”
Try the steps below if a 3Shape computer cannot access the shared server folders (even the server can’t access them).
Try this after checking windows firewall and antivirus for blocks.
1. Go to “Control Panel\Network and Internet\Network Connections”.
2. Right-Click on the connection and choose Properties.
3. Ensure that “Client for Microsoft Network” is checked.
4. Then click OK!
This should restore your ability to access the folder location. If not, call Technical Support.
How can I fix my PC’s not showing on the network (3Shape Issue)
The network connects all of the PC’s together to send their information on to the Printer or the Mill. If you do not see your PC listed on the network, here is the procedure to get up and going.
1. Go to Control Panel > Administrative Tools > Services. Then make sure that the Status field displays as “Running” and the Startup Type is set to “Automatic” for all of the following services;
DNS Client
Function Discovery Provider Host
Function Discovery Resource Publication
SSDP Discovery
UPnP Device Host
2. If they are not displayed as Automatic, double click on the applicable service and change the Startup Type to “Automatic”. If the Service status displays “Stopped,” click the “Start” button. Click “Apply” and then “OK”.
3. Now close out of all of the boxes. The PC should appear on the network and see other PC’s on the network.
How can I get 3Shape lab scans in an STL format?
! NOTE: ‘ScanIt Restoration STL Output’ option must be enabled on the dongle before scanning or exporting in STL format.
ScanItDental
It is not possible to scan directly in .stl format using ScanItDental, but the scans can be exported in .stl.
1. Scan the order in ScanItDental
2. In Dental Manager right-click on the scanned order
3. Go to Advanced > Export scan
4. In “Save as type” choose STL File (.stl)
5. Choose where to save the files and press “Save”
ScanItRestoration
It is possible to scan directly in .stl format using ScanItRestoration
1. Create an order and open it in ScanItRestoration
2. Before starting to scan go to File > Settings
3. In “Save type” select stl, press OK and start scanning
How do I Copy options in 3Shape Dental Manager work?
There are several options of copying an order in Dental Manager. All of them can be accessed by right clicking on the order and going to Copy:
Standard copy
This option will create a copy of the order with all the files and settings of the original order. If the original order was in the ‘Designed’ state its copy will be in ‘Scanned’ state, but as the DentalDesignerModellingTree.3ml is copied it is possibly to recreate the original design.
Copy and reuse CAD design
With this option it is possible to reuse design of the previously created Digital Temporaries and Diagnostic Wax-ups orders following these steps:
Right-click on the designed order and choose Copy > Copy and Reuse CAD Design;
A copy of an order is created with the Waxup/Pre-preparation scan enabled in the Order form.
Scan the new order or import a scan of the preparation (during scanning of the order SID will ask to align the new Preparation scan to the old one),
When designing a new order to select Choosing smile library > Provider > Reuse original teeth on the SmileComposer step to apply scans of the original teeth to the created crowns and pontics.
See also: Dental System user manual > How Do I Create > Temporaries > Reuse Design
Copy and Append CAD Design to Preparation scan
This option will attach the design that you made to the scan, and will create a new order where you will have original scan with appended design for the user to create a new design on top of it.
Append Design for Reuse
This option is used with abutment cases only. When a new scan is made, the system of coordinates of implants remains the same, and the design can be reused following these steps:
Create and design an order #1 with a customized abutment (doesn’t matter if it has anything on top);
Create order #2 with a customized abutment in the same place as in order #1 (doesn’t matter if it has anything on top);
Right-click on the order #2, go to Copy > Append Design for reuse and select order #1 in the list of the orders;
When designing order #2 on the Parametric design step you will have an option of using the of using the design made in in order #1.
How do I delete old orders from the 3Shape database?
This guide will walk you through the steps to setup the automatic order deletion function in 3Shape.
1. Open the Dental System Control Panel.
2. Click the General option under the Dental Manager heading.
3. Scroll to the bottom of the window.
4. Enable the automatically delete orders option.
5. Set the number of days that you want to keep orders. Example – Settings 30 days will delete all orders except those that were made within 30 days of today.
6. Check the box next to “Keep files but delete from database”. This will keep the order data on the computer but remove the entry from the database. If this is not checked, it will delete the order from the computer and there will be no way to recover the order if needed.
7. Click the Delete Now button.
8. Save and close the Control Panel.
How do I Import DME files into Dental System?
To import materials, libraries or settings of an external manufacturer into the Dental System the following steps must be performed:
1. Request the .dme file for the materials, libraries, or settings from the manufacturer.
2. Open Dental System Control Panel > Tools > Import/Export.
3. Select the Import Materials tab.
4. Select the .dme file from its file location.
5. Import and wait untill the materials are uploaded.
6. Once the import is finished, a confirmation message will show up.
Imported materials, libraries and settings will be shown in the respective sections in Dental System Control Panel.
! NOTE: Be aware that some materials are not approved in certain countries due to the FDA regulations, so they cannot be imported into Dental System.
How do I use 3Shape ‘USBview’ Application?
USBview is an application that enables you to browse all USB host controllers and connected USB devices on your computer. 3Shape has its own version (3Shape USBview) and is installed automatically together with ScanServer or ScanSuite.
3Shape USBview will help you connect the scanner according to the USB connection requirements.
3Shape USBview is located in C:\Program Files \3Shape\USB Test or C:\Program Files (x86)\3Shape\USB Test depending on a software version installed.
3Shape USBview is updating the status of USB connection automatically – each time the device is connected, disconnected or appears in another USB port, the change is displayed immediately – no need to refresh the window.
3Shape USBview user interface
1. USB Host Controllers (left panel).
Shows available USB host controllers in the PC
Shows version of USB used in each host controller (e.g. USB 2 = USB 2.0)
Shows if any of the devices are connected into the ports of the host controller;
Highlights the host controller with certain colors to show what devices are connected (e.g. green host controller color means only 3shape scanner is connected into it)
Each USB host controller is clickable to see which ports does it have and what devices are connected.
2. USB ports in the selected host controller (right panel).
Shows available USB ports in the selected USB host controller
Shows extended information about type of USB host controller in the right top part of the application window
Shows all devices connected into USB ports of the host controller
Highlights USB ports with certain color according to the type of connected device (e.g. green color means 3Shape scanner, blue means other device)
Each USB port is clickable to see extended information about the connected device.
3. Connected devices details (bottom panel)
Shows information about the device in the selected USB port, provided by the device manufacturer
For 3Shape lab scanners it is possible to see scanner model, serial number and USB driver version.
How to integrate Implant Studio in Dental System
To integrate Implant Studio with Dental System check the following,
Dongle(s) with active Dental System and Implant Studio license is(are) connected.
Dental System software is installed.
- Install Implant Studio (to integrate Implant Studio in Dental System standalone installation of Implant Studio is required)
- Open Dental System Control Panel > System Settings > click browse next to ImplantStudio path in the corresponding section. Browse / insert the following path,
If Implant Studio is installed on the same PC – C:\Program Files\3Shape\Dental Desktop\Plugins\ThreeShape.Triart\ImplantStudioIntegrated
If Implant Studio is installed on another PC in the local network: \\{ Computer Name Or Ip Address }\3Shape Dental System\Dental Desktop\Plugins\ThreeShape.Triart\ImplantStudioIntegrated
3. Save the changes in control panel menu.
Now you can create an Implant planning order in Dental System, upload or create scans and open the implant planning step using Implant Studio integrated.
I’m trying to install a server-client system, but I can’t find how to do it (2.17.2.1 Stand Alone upgrade to Server-client connection)
Steps to reproduce:
1. Install standalone DS;
2. Run installer on top and choose Server install;
3. On another PC try to install a Network Client to this Server installation.
Achieved result:
Not possible to set DS Server – “No Dental System found on” when setting PC name or IP address.
Expected result:
DS Server should be detected so that it’s possible to install Client.
Note: Client installation works fine when Server is installed clean, and not on top of standalone.
WORKAROUND: on Client PC go to shared folders of the Server PC and manually create DM and DSCP shortcuts. Manually Change the application paths pointing to C drive to the network paths. This fix will work best if you are in contact with the Technical Support Team when attempting.
Is there a way to import multiple patient records into OrthoAnalyzer?
You can if you exported all the cases to one file. See the following steps:
Export:
1. Select the cases you want to export by holding “Ctrl” on the keyboard and left clicking the case in OrthoAnalyzer.
2. With all the cases selected, left click the export button in the bar at the top left and select “export to file.”
3. Select “No” in the pop up box asking if you want to put all patients into a common file.
4. Select where you would like to save the file and name it. Click “Save.”
5. In the pop up box, select all the files you would like to export and click “Ok.”
Import:
1. In the bar at the top left of OrthoAnalyzer, select “Import from File.”
2. In the Open box that pops up, navigate to the file and select “Open.”
3. In the pop up box, make sure all files are select and select “Ok.”
The F1 Key is Not Bringing up the 3Shape Manual
This response just came back from 3Shape regarding an issue with F1 key not bringing up the user manual:
3Shape says it is bug #42179.
This should be resolved on the 2018.1.1 version of Dental System.
Unfortunately, 3Shape does not have the release date for that version just yet.
For the moment, the workaround is to open the manual directly from its path, which is C:\ProgramData\3Shape\DentalSystem\Documentation\2018\English\Manual
What ModelBuilder license packages are used in Dental System 2018?
In DentalSystem 2018 there are the following ModelBuilder license packages:
1.ModelBuilder All scans (MB STL + MB 3rd party + MB TRIOS + MB Impression licenses);
2.ModelBuilder TRIOS scans;
3.ModelBuilder Impression scans (sold together with ScanIt Impression as a part of Impression Scanning license package).
How does DentalDesigner 2018 work:
Any type of scan (including TRIOS, impression, STL, 3rd party) can be opened with DD Premium, Crown & Bridge or Complete Restorative subscriptions. ModelBuilder licenses are not checked in DD2018.
How does ModelBuilder 2018 work:
1. To work with TRIOS scans it requires ModelBuilder TRIOS scans license or 6 CAD Points.
2. To work with Impression scans made with a 3Shape lab scanner, it requires ModelBuilder Impression scans license or 6 CAD Points.
3. To work with STL scans and 3rd party orders it requires MolderBuilder 3rd party scans license or 12 CAD Points.
What ports should be open when using 3Shape Dental System?
When installing and using Dental System, the following ports should be open,
MS SQL Server: 1433, 1434, 1066, 1666
Dongle Server Service: 27027
3Shape Communicate: 9200
FTP: 21
HTTP: 80
SMTP: 25
If any of the ports are not open, the user should contact their local IT assistance for opening the ports.
Where do I find Dental System application logs?
Logs for Dental System applications can be found under following locations,
Dental System Installer – %ProgramData%\3Shape\DentalSystemInstaller.log
Dental System Control Panel – C:\Program Files\3Shape\Dental System Control Panel\ControlPanel.exe.log
Dental Manager – %ProgramData%\3Shape\DentalManager\DentalManager.html
Dental Manager Service – C:\Program Files\3Shape\DentalManagerService\DentalManagerService.exe.log
Dental Designer – C:\Program Files\3Shape\DentalDesigner\DentalDesigner.exe.log
3Shape Communicate Inbox – %ProgramData%\3Shape\DentalManager\ThreeShapeCommunicateLog.html
ScanItDental – %ProgramData%\3Shape\ThreeShape.ScanItDental Standalone\ExceptionLog.txt
ScanServer/ScanSuite – %appdata%\3Shape\ScanServer\ExceptionLog.txt
Dongle Server Service – C:\Program Files (x86)\3Shape\Dongle Server Service\DongleServer.exe.log
Why isn’t my Crown and Bridge Dongle is not recognized by the 3Shape Software?
For the crown and bridge dongle, the 3shape version must be 16.2.0 and newer for the dongle to be recognized.
If you are having trouble with recognition, call or e-mail Technical Support for up-grading your 3Shape version to the 16.2.0 Version or newer.
Will 3Shape automatically out-put my Manufacturing Files?
3Shape can automatically output design files for local milling. However, this function is not enabled by default. Please take the following steps to set up automatic out-put.
1. Open the 3Shape Control Panel
2. Go to Manufacturing Processes
3. Change the user’s materials to only have their site as the manufacturer.
You need to do this for each manufacturing process tied to a material that the customer uses for local manufacturing.
How do you import material ‘ini’ files?
Objective: The main purpose of updating replacing customer material files in Asiga is to ensure the customer is using the proper and most up-to-date print strategy. Print strategies can change due to advancements in the firmware, software, and even new printer classes.
How to delete out old material files- There are two ways to do so.
Method 1
1. Open composer and select new build (if composer is already open- locate the build properties icon and select it).
2. Under the “Material” section select the material you would like to remove and then select on the gear.
3. The gear will drop down with multiple options- select on the delete option if available. If the delete option is greyed out, try method 2.
Method 2
1. Open file explore and browse to the following path to locate saved materials
C:\Program Files\Asiga\Composer\Materials
2. Select and delete the material files that you are wanting to remove from the Asiga Composer Software.
How to import new materials into the composer software
1. Transfer the new material files to the customers PC either by email or by another files transfer service (example: TeamViewer, Wetransfer, etc.).
2. Open the Composer software.
3. Select “New” for creating a new print job- the Build Parameters window will open.
4. Select on the “Gear” on the right next to the material selection drop down.
5. After selecting the “Gear”, click on the “import” option. If using Composer 1.2.8 you will then select “composer”.
6. From here, a window browser will open and you will need to locate the material file that you would like to import. If you downloaded the material file from an email or link there is a good chance that the file ended up in the “Downloads” folder. After finding the downloaded material file select it and then click “open” in the bottom right corner of the window browser.
7. Verify that your material file is loaded into the software by searching for it in the material selection drop down. If it is there, you have successfully imported a material file.
8. Enjoy!
How do I repair the “Trios Inbox Standalone Error: TRIOS Scan Data: Missing node Add Order Request” error?
The information from 3Shape is as follows:
1. Explore Order
2. Order Source
3. Erase all files with 1kb
It would be best if you let the Technical Support Team walk you through this issue.
What can I do if a Doctor sends a scan of an analog impression for model printing?
Unfortunately, inform the customer that this is not a valid workflow.
The Trios, is for scanning positive objects such as teeth in the mouth or a stone model. If a scan of a negative surface is done and sent in for processing, the Dental System will treat it as if it is inside out and the Model will be built in the reverse direction. In other words, you can build a model of the impression but you can not build a model of the reverse of the impression.
Are the images stored on the iPad or iPhone?
No, all facial images are stored in the Bellus3D cloud.
Are there additional fees besides the annual subscription?
No, the annual subscription covers all imaging, file sharing, and exporting.
Can I scan some shoulder part?
There isn’t a mode for capturing the shoulders at this time.
Can you use the app with any iPad?
No, it must be the iPad Pro (without the thumb pad) to have access to the high tech camera. The app also works with iPhone 10x or 11x although it is hard to adjust with a small screen.
How can I delete a scan I already made?
Here is how to delete a face model or workspace.
- From the gallery, upper left corner of iPad, press Edit
- Touch the face model(s) you would like to delete
- The face model will have a check mark in the upper left corner
- Upper left corner of iPad, press Delete
- Face model is deleted
How can I learn more about this app?
Go to the administration toolbox icon and click on either Feedback & Support for the Knowledge Base or Tutorial for taking 3D images.
How do you bring scanned models into the app?
If you are using 3Shape, login to your 3Shape account in Bellus3D, and you will have access to your files. If you are using other CAD software, have the scanned models in cloud software such as google drive, and bring in the files from that location.
How many 2D photos are taken with the app?
Seven 2D photos are saved with the 3D image.
Is there any difference in resolution when we export in OBJ or STL file?
No, but the OBJ will have color/texture data.
What are Orthographic Projection images?
These images relate to perspective. When you are looking at a distance, images are smaller. If you want to measure accurately, use the Orthographic Projection view to maintain relationship size. Find more information in Feedback & Support, Knowledge Base (in Settings).
What is the difference between the free version of Bellus3D and Dental Pro?
The free version of Bellus3D allows for facial scanning only. You cannot make any changes to the images and there is a cost to export the file.
Where can we purchase the light ring similar to the display version?
SOCIALITE 18" LED iPad Ring Light Kit – Incl. Light, 6ft Stand, iPhone/iPad/DSLR Mount, & Remote
You will also need a portrait iPad holder:
Will the app notify me when the subscription is about to expire?
Not at this time. That is in development for a future upgrade. At Whip Mix we strive for superior customer service and we will notify you prior to expiration.
Will the dentist have any access to tooth libraries within Bellus3D?
Coming soon, the dentist will have access to dental arches. Pick between different tooth shapes and styles, and arch sizes for a general idea of what will work for the patient.
Are there patient instructions for the home use of the Nox T3?
Yes. Contact Whip Mix Technical Support for a duplicate copy.
Can a dentist use his or her own scan reading network?
Yes the dentist can. It is advisable that they work with a local Physician if possible so that there is more of a connection with the Patient, Dentist and Physician.
Does Whip Mix have a Network for reading the sleep scans? Who are they?
Whip Mix has relationships with Medical Doctors in each state. We do not give this information out so that we can protect our connections. There is a cost and a code to access. Contact Whip Mix Technical Support for further information.
Is there Technical Support for Gem and Gem Pro?
Vyaire provides Technical Support 24/7. The number is 1-800-231-2466
Call Technical Support if the monitor stops recording.
What are Practitioners charging for a sleep test?
Generally between $0 and $250.
What are the Device Definitions for Sleep Testing Monitors?
What are the Farangometer and the Rhinometer?
Who is Apnea Link?
APNEA Link is a competitive home sleep study. Medical doctors like to read the type of data and software they are used to reading…so if a dentist has an existing relationship with a Medical doctor, he/she may dictate what unit the dentist buys. You should be aware of this brand.
How do I convert my Whip Mix, Denar or Hanau Articulator to Magnetics?
Why should I use an articulator?
The articulator is a mechanical device that simulates jaw movement. Unmounted casts give the technician only basic information about the patient’s occlusal relationship. For a complete analysis of the functional relationships, the use of an articulator is required.
What is the difference between an arcon and non-arcon Articulator?
In an ARCON designed articulator condylar housings are built into the upper member of the articulator and the condyles are always fixed onto the lower member of the articulator. This design simulates the location of the condylar ball for the patient. With a non-arcon designed articulator, the condylar housing is found on the lower member of the articulator and the condyles are attached to the upper member. This design is opposite of the anatomy for the human skull.
What are the advantages of using a Semi-Adjustable articulator?
The use of Semi-Adjustable Articulators, like the Whip Mix, is having the capability of providing the maximum amount of diagnostic information with minimum need for clinical adjustment during delivery/placement of the appliance.
[Video] Why Do We Need Bite Records, Facebow Transfers & Semi-Adjustable Articulators?
When would I use an Immediate Side Shift (I.S.S.) setting?
In some patients the timing of the Bennett movement occurs early in the latero-trusive movement. When this occurs a shift is seen even before the condyle begins its translation from the fossa. This is called Immediate Side Shift. Accommodating the patient’s side shift movement will offer more freedom of lateral movement while the patient is in centric relation. It is important that the laboratory technician program the articulator to ensure the movement replicates the patient’s condition. The Whip Mix Model #2240 offers the Immediate Side Shift capability.
Do I need to use a face-bow to use an articulator?
No you do not however, it would be best to provide as much detailed, patient information as possible to reproduce the patient’s ideal occlusion.
A 90 Second Facebow Can Save You Time & Money in Your Dental Office
[Video] Why Do We Need Bite Records, Facebow Transfers & Semi-Adjustable Articulators?
Can I convert my articulator over to the magnetic system? If so how?
Many Semi-Adjustable Articulators on the market can be retrofitted to use the Quick Mount Magnetic System. Refer to the selection chart on the Quick Mount Magnetic System by going to the link above to determine if your articulator can be converted.
Do you have an accessory that allows me to make a custom incisal guide table?
Yes, the Dovetail Incisal Block. During the wax-up procedure on the mounted stone casts, the laboratory technician or dentist may have to move the casts through frequent excursive movements on the articulator to check the positioning of the wax to ensure ideal occlusion is achieved. During this procedure wearing away of the stone casts is likely. To prevent this unwanted wear, a custom incisal guide table made out of auto polymerizing acrylic resin should be fabricated either on the flat plastic incisal guide table that comes with the articulator or on our metal Dovetail Incisal Block.
Is there a way to quickly mount study casts to on an articulator without mounting them with plaster?
Yes, our InstaMount Mechanical Mounting System, a plasterless mounting system, allows you to mount your study casts on your semi-adjustable articulator. This is simply achieved by placing and securing your upper and lower casts on the Model Platform. Next, tighten the platform to the InstaMount Base Plate. After you have checked for and eliminated possible occlusal interferences, bring the casts into occlusion. While holding the casts together center them in the middle of the articulator and hand tighten the locking screw. After final adjustments have been made, firmly lock the InstaMount in place using the locking wrench.
What is the 8500 and is it FULLY Adjustable?
This is the original Whip Mix Articulator, which has adjustable condyles, and is based on the Stuart Articulator. Does this make it a “Fully” adjustable articulator?
While you can adjust the condyles from Small to Medium to Large (S, M, L) as noted on the face of the lower member of the articulator, just under the condyle, and indicated when using the Whip Mix 8645 Face Bow; these are still specific spots in movement. The condyles are then not “Fully” adjustable which would be necessary for it to fall into that category. It remains a Semi-adjustable Articulator.
Why should I use a Whip Mix Articulator?
Simple! The Whip Mix Articulator System is extremely easy to use. It is not intimidating for dentists who may be thinking about using a Semi-Adjustable Articulator. The simplicity and speed with which the necessary registrations are obtained and transferred to the Whip Mix Articulator (using the Whip Mix Face-Bow) enable the operator to accomplish corrective and restorative dentistry with greater precision than ever before; and without the use of expensive equipment and more time consuming techniques.
Affordable! The Whip Mix Semi-Adjustable Articulator and Face-Bow System is one of the most affordable and reliable systems on the market.
Durable! Utilizing parts made of anodized aluminum, phosphor bronze or stainless steel, assures a maintenance-free use. With over 40 years of manufacturing experience, the Whip Mix Articulator has established itself as an instrument of choice for the majority of undergraduate dental education programs in the US and Canada.
Interchangeable! The Whip Mix 2000, 3000, and 4000 Series Articulators have the “ACCUMOUNT” Mounting System. This makes it possible to interchange mounted casts between individual series of articulators without the loss of accuracy. The accuracy of interchangeability is reported in the Journal of Prosthodontics, vol 9, No 3 (September), 2000.
Read More – Answers to 5 Frequently Asked Questions About Whip Mix Articulators
[Video] Why Do We Need Bite Records, Facebow Transfers & Semi-Adjustable Articulators?
What type of design are Whip Mix Articulators?
All Whip Mix Semi-Adjustable Articulators feature the ARCON design. ARCON is a contraction of the words for articulator and condyle. In an ARCON designed articulator the mechanical fossa is fixed relative to the maxillary cast, making the instrument more anatomically correct. Because of their ARCON design, Whip Mix Semi-Adjustable Articulators are ideal for the study of occlusion and the movements of the temporomandibular joint.
What features do Whip Mix Articulators possess, and how do I know which articulator model is right for me?
This will depend on what you need your articulator to do. Whip Mix offers several models that offer different features depending on the extensive requirements for the patient being treated, needs of the dental laboratory, or the needs of the dentist.
The Model #8500 Articulator, an open fossa design, features a straight tracking condylar element, progressive side shift setting, and adjustable intercondylar width. The intercondylar width can be set at the following distances: 96mm (small), 110mm (medium), and 124mm (large). This distance is determined by the marks found on the face-bow during the face-bow registration procedure. This articulator is not interchangeable.
The Model #2240 Articulator, an open fossa design, features a fixed intercondylar width of 110mm, curved emenitia with a ¾” radius, fixed bennett angle set at 7½ degrees, has immediate side shift capability and is interchangeable within the 2000 series.
The Model #2340 Articulator, an open fossa design, features a fixed intercondylar width of 110mm, straight condylar element with progressive side shift capability and is interchangeable within the 2000 series.
The Model #3040 Articulator, a closed tracking design, features a fixed intercondylar width of 110mm, progressive side shift capability and is interchangeable within the 3000 series.
The Model #3140 Articulator, has a modified tracking design, features a fixed intercondylar width of 110mm, progressive side shift capability and is interchangeable within the 3000 series.
The Model #4640 Articulator, offers the most visibility of any Whip Mix Articulator available on the market. This is made possible by a design that eliminated the crossbar between the upright posts. The Model #4640 Articulator features a closed tracking design, a fixed intercondylar width of 110mm, progressive side shift capability and is interchangeable within the 4000 series.
The Model #4641 Articulator, like the Model #4640, offers the most visibility of any Whip Mix Articulator available on the market. The Model #4641 Articulator has a modified tracking design, a fixed intercondylar width of 110mm, progressive side shift capability and is interchangeable within the 4000 series.
All Whip Mix Semi-Adjustable Articulators are also available with the Quick Mount Magnet System.
The Model 100 Simple Articulator features a magnet mounting system, fixed protrusive angle at 20 degrees, fixed lateral setting at 15 degrees, reusable metal mounting disks, adjustable mounting platform stand, and flat and curved mounting platforms. The Model 100 Simple Articulator is recommended for the less complicated cases. The Model 100 Simple Articulator is not interchangeable and does not accept a face-bow.
Are Whip Mix Articulators interchangeable and why is having an articulator that is interchangeable important to me?
Yes. With the exception of the original Model #8500 Articulator, all Whip Mix Semi-Adjustable Articulators feature the “ACCUMOUNT” system of interchangeability which makes them interchangeable within like series. The accuracy of the interchangeability is reported in the Journal of Prosthodontics, vol 9, No 3 (September), 2000.
Cost savings. There is no need to purchase several articulators for your practice/laboratory. Using an articulator that is interchangeable allows you to exchange mounted casts between like articulators without loss of accuracy. Only the mounted casts need to be sent between the doctor’s office and the laboratory. There is an additional savings in shipping cost for the weight of the articulator being sent back and forth between the doctor’s office and the laboratory. Finally you would eliminate repairs that are a result of the articulator being damage in transit between the office and laboratory.
How do I know if my Whip Mix Articulator is interchangeable?
Any articulator with the “ACCUMOUNT” plate offers interchangeability. This can be identified by a 2 ½ ” circular plate that is attached to the lower member of the articulator.
Are all Whip Mix Articulator models interchangeable with each other?
Yes, the Whip Mix 2000, 3000, and 4000 Series Articulators are modified with the “ACCUMOUNT” Mounting System. This makes it possible to interchange mounted casts between like series of articulators without the loss of accuracy. The accuracy of the interchangeability is reported in the Journal of Prosthodontics, vol 9, No 3 (September), 2000.
Do I have to program my Whip Mix Articulator on each case or can I just use average settings?
Average settings will accommodate most treatment cases. Although, you do have the convenience of customizing your settings if you have taken lateral check bites. Whip Mix uses average settings of a 15° Bennett Angle and a 30° Protrusive incline; or condylar inclination.
I have an old Whip Mix 8500 Articulator, what do the newer models such as the 2000, 3000, and 4000 Series have to offer?
The most significant difference between the Model 8500 and all other models is the interchangeability feature. In addition to being interchangeable, the 2000, 3000, and 4000 Series Articulators offer increased interframe distance and more lingual visibility. For more detailed information on each series, please select from the appropriate link.
What accessories are available for the Whip Mix Articulator?
Whip Mix accessories enhance the capabilities of the Whip Mix Articulator systems and provide a wide range of options for specialized needs. For a complete description of available accessories go the products page then click on the articulator section. Once you have selected the articulator section click on the accessories.
I’m looking for an economical articulator to use for my simpler cases. What would you suggest?
In many instances you may not need the use of a Semi-Adjustable Articulator for the less complicated cases. A simple articulator with a fixed condylar inclination and lateral settings will work just fine. Whip Mix’s Model 100 Simple Articulator is the answer. This articulator features an easy-to-use mounting system with indexed magnetic mounting plates and reusable metal mounting disks. Average settings allow for smooth protrusive and lateral movements. A Mounting Base with both Curved and Flat Mounting Platforms and an Anterior Centering Pin enables the upper cast to be quickly positioned for easy case mounting. The lightweight aluminum frame incorporates a retractable anterior leg and a posterior stop for optimum tabletop viewing.
What are the advantages of using the Quick Mount Magnetic Mounting System?
The most notable result is saving time and money. The Quickmount Magnetic System allows the quick, precise and easy placement and removal of mounted casts on many of the most popular semi-adjustable and fully adjustable articulators.
How can I tell if my Whip Mix Articulator is out of alignment?
If my Whip Mix Articulator is out of alignment, what do I need to do to have it fixed?
If you feel your articulator has been damaged please call our customer service line at 800-626-5651. Our Customer Service Representatives will be happy to assist you with our repair process. Have the model and serial number available with a brief description of the problem ready to give to the Customer Service Representative. You will be issued a Return Authorization Number (RMA) which will be used to track your articulator when it arrives.
Could I use a Mark 320 articulator for a full mouth rehabilitation, would it be acceptable instrument to complete a complex case?
I have seen many complex cases done with less of an instrument than this… The 330/320 would certainly handle any of the complex cases including full mouth reconstructions where you have the opportunity to correct some of the outlying issues with the patient. The ability of capturing Arc of Closure, Anterior envelope of function, Wilson and Bennet are key. Until you have Video CT and are able to use it directly in the computer with design software for creation of the restorations, a mechanical instrument like this is your best option.
How can I replace the latch spring for the D5A?
Recently a doctor contacted me looking for parts for her older Denar articulators. As we went over the parts she needed one of the parts she needed was the latch spring for the D5A (part # 110157-1). The spring had broken and needed replacement.
After further discussion, we discovered she needed a spring for another instrument over 20 years old, the DG100 (part of the Gnathomatic System). I was unsure if even something as simple as a latch spring was available in this case. As is normal I contacted our helpful team at Whip Mix West, and as luck would have it the # 110157-1 actually worked.
How does the Mark 330 accept an immediate side shift? Where on the articulator are the adjustments located for this measurement? Do you need to have a Cadiax to record an Immediate Side Shift?
The immediate side shift adjustment is made using the same knob used to adjust the progressive side shift. The calibration markings for this measurement are directly behind the knob. There is a 0-2mm adjustment. Cadiax is one of the better instruments in determining not only immediate side shift but also the progressive side shift and condylar inclination as well. However, there are other techniques for determining these adjustments in other philosophies. Ultimately, adjustment in the patients’ mouth is the best articulator.
I understand the protrusive record gives you the condylar inclination, lateral record gives you progressive side shift, etc. If the patient did have an ISS and I did not realize that they did, would there be any indication again such as the examples that I mentioned above.
If the patient is beyond 15 degrees in progressive side shift, how would I realize this on the Mark 320/330 articulator?
The Denar Mark 300 series was designed to have “patient averages” built into the design. The short answer is no you would not be able to set the Bennett angle past the 15-degree mark built into the design.
The design of this instrument and many other instruments manufactured by Whip Mix as well as many other articulator manufacturers took the normal practice used by most general dentists. By this we mean most general dentist, do not take full protrusive and lateral records to set up the articulator. In this case having patient averages built in made it more convenient for both dentist and technicians using an articulator as part of the armamentarium for manufacturing a prosthetic appliance. The literature tells us that most patient’s actual Bennett angle falls somewhere around 7-12 degrees. Therefore, if the doctor doesn’t provide a technician with actual a true angle setting, then setting it to 15 degrees will ensure that interferences during lateral movement will be minimized or in most cases completely avoided because the 15 degree setting opens up the angle of travel for opposing tooth surfaces.
Is the Mark 320 a curvilinear or rectilinear articulator?
The Denar Mark 300 Series has a rectilinear or straight lined superior wall.
Since the Mark 320 cannot accept an ISS record does it then already have an ISS programmed in? If so what is it, is it based on averages, would it be set at 0.5, 0.75 or 1 mm already?
The Denar Mark 320 series was designed to have what are called “patient averages” built into the design. With this design of the Mark 320, ISS is not available nor is it built in.
The thumbscrews that hold the mounting plates to the articulator on my Combi are worn out. Can I replace them?
Thumbscrews for the COMBI and the COMBI II are the same part (ITEM# 110083) even though they are noted with different engineering numbers in the drawings.
If you pull slightly up on the upper or slightly down on the lower thumbscrew, you will be able to unscrew the thumbscrew from the articulator.
The thumbscrews are held in place by a THREADED BUSHING (ITEM# 113033), both in the upper and lower members of the articulators. If these bushings are damaged, the articulator must come back to the factory for bushing replacement. The bushings are pressed into and out of place by a piece of machinery, which is not available in the lab or clinic.
What type of record would you need to take for determining ISS on the Mark 330?
ISS can be captured only by taking a digital or mechanical tracing. You cannot just simply ask the patient to move their jaw around and from there make a determination if ISS is present and if so how much. There are products available like the digital pantograph or our own Cadiax Compact 2, which are capable of determining if ISS exist. Some well-known and highly respected clinicians believe that if you seat the patient’s condyle into their fossa, you will always have a stable and repeatable position from which to work. This of course is Centric Relation position, (CR).
Why does the Mark 330 only give you an adjustable progressive side shift range of 0-15 degrees?
This is a Semi-Adjustable Articulator. When we first started the design of the Mark 330 series, Whip Mix reached out to some Dentist connections, which included Dr. Peter Dawson, Tuffs University Faculty and others to help determine the criteria for this articulator. It was determined at that time to go with 0-15 degree progressive side shift as an average that would cover the majority of patients. If the patient were beyond the 15 degrees, they would need something more like the D5A fully adjustable articulator. When you are working on Dentures, you generally use a progressive angle of around 7 degrees. The Mark 330 makes this possible.
Would there be for example an obvious gap in the condyle fossa of the articulator or some kind of occlusal discrepancy that would be visible on the mounted cast, etc.??
Would you use the same lateral inter-occlusal record that you took for progressive on the Mark 330?
The terms Lateral record and Progressive record has the same meaning in many dental circles. Lateral means that the jaw is sliding laterally, (to the left and right). Progressive means the degree of angle the jaw is moving as it slides from left to right. So if this record is being used to find the angle the condyle is moving down the eminence then use you can use this record.
Do you have parts for my Hanau Articulator?
If it is not a current production model of a Hanau, we most likely do not have parts.
Please take a photo showing the entire Articulator, then another showing the condyles, and another from the back of the articulator and then contact Technical Support.
We can help you find out if there are parts available or suggest that you mount a nice set of models to the articulator and use it in your office as a very nice piece of history.
What is the History of Hanau?
Few names in dentistry have been associated with articulator design and manufacture as much as that of Hanau. Indeed, over the past 85 years, many have considered the name Hanau to be synonymous with the word articulator.
Rudolph L. Hanau was born in Victoria West, Cape Colony, South Africa in 1881. He was trained as a mechanical engineer in Leipzig, Germany. By 1906, he had immigrated to the United States where he worked as a consulting engineer in New York City.
Dr. Frederick Lester Stanton, a New York City orthodontist, conferred with Hanau in 1915 to assist in devising a practical method to determine dental arch form. Their association led to the development of the Stanton-Hanau Surveyor (Fig 2).
This instrument was intended for orthodontic research and was never meant to map the contours of teeth. No evidence has been found to suggest that this surveyor was ever sold commercially.
Hanau left New York City in 1917 to live in Pittsburgh, PA. He again relocated in 1918 to Buffalo, NY. In 1920, the legendary Rupert E. Hall consulted with Hanau concerning a new articulator design. This challenge set the course for the remainder of Hanau’s professional career. Hall’s intent was to develop an adjustable three-dimensional instrument with a central vertical axis, around which the condyles orbited during function (Fig 3).
He wanted Hanau to build the prototype. In the end, Hanau chose to follow the conventional approaches of the time. His concept and resulting landmark Model H evolved from the designs of Gritman, Gysi, Snow, and Wadsworth.
Hall’s consultation with Hanau may have ignited the engineer’s passion for articulator design; however, Hanau’s epiphany came during one of the heated battles of the early 20th century “Articulator Wars.” In 1920, Hall invited Hanau to attend the Boston meeting of the National Society of Denture Prosthetists. Many of dentistry’s “Titans” of the day were in attendance. Hanau was exposed to the likes of Drs. Alfred Gysi, George S. Monson, and his host, Rupert E. Hall. Each presented papers on his personal theory of mandibular movement and articulator design. As an engineer, Hanau perceived the problem of articulation quite contrary to what he heard at the meeting. Upon returning from Boston, Hanau launched the full effort of his engineering department to design and develop a prototype articulator. In just a few months, he had produced his Model Articulator. He applied for a US patent in February 1921 (Fig 4).
Within another month he improved his design, which became the Model B articulator (Fig 5). The Model C was Hanau’s final refinement in this series of instruments. He revealed it at the Milwaukee meeting of the National Society of Denture Prosthetists in August 1921. The Model C was commercially produced under the Model A patent, which was eventually granted on June 1, 1926.
Journal of Prosthodontics 19 (2010) 409–418 c 2010 by The American College of Prosthodontists
Which Mounting Plates do I use for a HANAU Articulator?
There are several options and these include the following:
ITEM# 005057-000 Metal Mounting Plates (these are the original metal, round, Zinc plates, 1 pair)
ITEM# 1100026-10 Denar/Hanau Disposable Plastic Mounting Plates (quantity 10)
ITEM# 1100026-50 Denar/Hanau Disposable Plastic Mounting Plates (quantity 50)
ITEM# 1100026-100 Denar / Hanau Disposable Mounting Plates (quantity 100)
TOS SYSTEM PLATES for Orthognathic Surgery Processes:
ITEM# 1100027-10 TOS Surgical Mounting Plates (quantity 10)
ITEM# 1100028-20 TOS Surgical Mounting Plates (quantity 20)
Magnetic Mounting System
Hanau borrows from the Denar system to let you use magnetics on your HANAU Articulators:
ITEM# 20002367 Magnetic Converter Plates (one pair) you must also purchase disposable Mounting Plates.
ITEM# 20002368 Disposable Magnetic Mounting Plates (20)
ITEM# 20002369 Disposable Magnetic Mounting Plates (100)
I have Cracking in my Mainstay Articulators. Why?
Recently a customer called concerned that his Mainstay Articulator plastic sections were cracking vertically after mounting. During our discussion, I went over the importance of placing the pins finger snug before stone is applied and using a ‘food grade’ silicone spray to release the set stone after mounting.
The phenomenon was sporadic and had been happening for several months. I informed the technician I had only heard of this with one other customer and it was many years ago. I also mentioned the Mainstay Articulators are used by technicians around the country and internationally and I had never heard o this issue as and industry wide problem.
Photos of the issue were emailed to Whip Mix, and as the customer described there were indeed vertical cracks in the articulator sections. We were in the process of bringing in samples to evaluate the issue when the customer called back and said he believed they found the cause of the cracking.
After talking with the technicians using the product, it was discovered that they were forcing the pins into the holes using ‘above and beyond’ finger pressure, in an effort to seat the pin completely.
I explained that while ideal for the collar of the pin to meet the base of the articulator section, it is not necessary from a functional aspect of the system. Any pin that may have an elevated collar will have the pin section buried in stone and not cause a mounting or clearance issue. The important aspect of the pin is that it is snug in the receiving hole of the articulator to prevent dies from rotating. The long taper of the pin will engage the receiver hole over a long distance. This allows the pin to be high and still engage properly.
What is the difference between a Direct and Indirect Mounting Face-Bow?
Direct Mounting simply means that after you have obtained your face-bow registration and have removed the ear-bow halves from the patient, you transfer the face-bow directly on to the articulator to mount the maxillary cast.
With Indirect Mounting, once you have obtained your face-bow registration, you remove the Transfer Assembly from the cross bar and place the assembly on to the articulator using the Transfer Base Assembly for mounting the maxillary cast.
What are the advantages of using an indirect mounting Face-Bow?
Indirect Mounting offers you increased stability when mounting, greater access to your cast, and parallel orientation of the upper and lower members of the articulator. Additionally, there is no need to have several face-bows for your practice. Having additional Transfer Assemblies will allow you to utilize your face-bow immediately after proper asepsis.
I think I have an older Direct Mounting Face-Bow; can I convert it to an Indirect Mounting Face-Bow?
Yes, converting your Direct Mounting Face-Bow to an Indirect Mounting Face-Bow is very easy. You will need the following Conversion Package. 1 Transfer Base Assembly (#9176A) or (#9176QA) if you have a magnetic mounting system. 1 Cross Bar (#8679) and 1 Transfer Assembly.
How can I prevent my Face-Bow fork from flexing while mounting the maxillary cast?
Many clinicians use a Face-Bow Fork Support while mounting the maxillary cast. The Face-Bow Fork Support attaches to the lower frame in place of the lower mounting plate. Its cross arm is raised to gently contact the under-surface of the face-bow fork to prevent flexing of the fork.
I have an older Whip Mix Face-Bow. Have you made any changes?
Improvements have been made to Whip Mix Face-Bows. Two design changes you will notice are: a new soft plastic nose piece and a fixed elastic band which provides spring to the ear-bow halves. Not only is the new soft plastic nose piece more comfortable to the patient, the contour of the nose piece fits the patients nose more accurately.
Our newest Face-Bow model is the #9155 with the Quick Lock Toggle. It is universal in that it adapts to any Whip Mix Face-Bow for direct or indirect mounting. It is easier to use than the other face-bow styles because of the single toggle, easy to place bite-fork, and universal joint.
Can Whip Mix Face-Bows be used with other articulator systems?
Yes, you can adapt the Whip Mix #9185 (metal) or the #9285 (Plastic) Indirect Mounting Face-Bow to the Hanau Wide-Vue Articulator using the #9176A Transfer Base Assembly and the #9186 Face-Bow Adaptor.
What accessories are available for the Whip Mix Face-Bow?
There are numerous accessories that can be purchased and used with the Quick Mount Face-Bows. For a complete list and description of their use, please refer to the following link: Face-Bow Accessories. For accessories specific to the Direct Mounting, Indirect Mounting, or Indirect Mount with Quick Lock Toggle, link directly to the product.
I need to buy parts for my Cadiax. How can I do that?
Do you still sell the Accuceph & Accurad?
Whip Mix no longer sells or supports these pieces of equipment.
Refer all questions to Goodex (formerly Imaging Systems) 1-800-628-1302
What is the RDA (Relative Dentin Abrasive Index) of Pumice Preppies?
A customer called and wanted to know the ‘RDA’ , Relative Dentin Abrasive index, of pumice preppies. Preppies are not rated the way that Toothpastes are rated in terms of abrasiveness. In Toothpaste, it is very important because the patient uses it every day. Preppies are used only occasionally to remove build up.
The pumice particles in the preppies are 90% less than 40 microns and breakdown under use to smaller pieces.
Can the Learn-A-Prep II be used for learning “Full Contour” preparations?
Unfortunately, no. These products are both comprised of three layers colored white, yellow and Pink and have a differentiating hardness to help develop tactile skills while preparing teeth for restorations. The preparations would be occlusal, buccal, lingual or interproximal but not 360 degrees as the layers are flat.
What is the difference between Learn-A-Prep II and Mannequin Learn-A-Prep II?
Mannequin Learn-A-Prep II is the same product, but with an attached base designed to be used inside a dental mannequin training head, for practice on performing preparations in the mouth. It can be used with any dental mannequin head that has a ¼” x 28” screw size.
Can I use any of the Special Liquids with my phosphate investment?
The answer is yes, but it is complicated.
We can tell you that all three of the liquids will work with the investment powders, but they will have different expansions than when you use the correct liquid. We have not tested these combinations and cannot tell you exactly what will happen and by how much.
The SLC Plus has the greatest percentage of colloidal silica and drives the greatest percentage of expansion, so investments that use regular SLC would have higher expansions than normal when using SLC Plus. On the other hand, Formula 1 which uses SLC Plus, would have lower expansion than normal if you used regular SLC liquid.
My special liquid is turning a black color…What is happening?
Your liquid has bacteria in it and the bacteria is growing and living on the colloidal silica. You need to throw that material out and replace it with new Whip Mix Liquid.
The liquid is colored now instead of clear. Does the color affect the liquid in any way?
No, the color does not affect the liquid. Special Liquid Concentrate is a light blue and Fastfire Liquid is a purple color. Special Liquid Concentrate Plus is clear.
What happens if the Special Liquid freezes?
If you have Special Liquid Concentrate (SLC) or Fastfire Liquid; let the liquid thaw, then shake vigorously to reincorporate. Use as normal.
If the liquid is Special Liquid Concentrate Plus, you will not be able to reincorporate the liquid after it thaws. Replace the liquid with new and dispose of the old liquid.
I am getting incomplete castings/pressings… What is going on?
Generally, temperature issues cause incomplete castings. I know we hear that “it was not burned out all the way”…but really what do you think would happen to left over wax when alloy at over 900 to 2000 degrees F touched it?
The true answer is that the Mould or the Alloy were not hot enough. Since we are looking at the alloy, and we can see that it has become molten…it is probably hot enough. We cannot see if the glass is hot enough to press, but the results tell us if it is.
So, let us look at the ring temperature. Let us assume…that the oven is actually at the temperature that the read out indicates. Although I would venture to say that, you do not know the last time that you actually calibrated the furnace.
The larger the mass of the thing you are trying to heat, the longer it takes to heat. Therefore, at a fixed temperature, a 300gr ring takes longer to heat than a 100gr ring. The same is true if you put four 100gr rings in the oven at the same time. The rule is that you add 10 minutes of burnout time for every additional ring. So if one ring requires 30 minutes of hold time; four rings would demand?? That is right, 60 minutes. No… you cannot just increase the temperature.
Rings do not automatically heat up at the same rate that the air in the oven heats up. The thermocouple (TC) is reading the temperature of the air, not the inside of the ring. The rings lag behind, so we hold them at the top temperature so that they can catch up temperature on the inside.
Example: you have a short, rounded casting on the margins, but the button is sharp edged on the outside.
The alloy was hot enough, but the internal of the mold was not. The mould pulled the heat from the alloy and stopped it in its tracks.
The same is true for pressed glass. Short rounded edges mean a cold mould. Not the need for more press pressure or more press time. Try increasing the hold time five minutes, then the temperature by five degrees and then the hold time another 5 minutes until you succeed.
Example: if the button has rounded edges, and you have rounded edges on your margins. Then the alloy was not near the required casting or pressing temperature. Re-think how you are delivering heat to your button or alloy.
Now, pull out the silver strips or ‘Tempill’s ‘, and check that burnout oven. You may find that it is way off. If it is right on the temperature, then you are going to have to change your process for casting and pressing to find success.
I used to order a marking pencil from Whip Mix. It was dark blue and then turned red when the investment burned out. Can I still get it?
Unfortunately, this item was discontinued several years ago.
However if you make a visit to a good Stationary Store or a Craft Store that sells PRISMACOLOR colored pencils, you may find it.
It is; Prismacolor Verithin 741 Bleu Indigo.
Can ResinVest be used with a Standard Burnout Procedure?
When needed, ResinVest may be burned out in a standard manner (slow technique) with some modifications of the rapid technique. You will want to adhere to the 22ml per 100gr powder ratio using whatever proportion of liquid to water that has shown a good result from previous castings.
A good starting point is 75-80% Special Liquid Concentrate Plus with the remainder distilled or deionized water to make 100%. This ratio would be for printed partial frameworks with two frames in a 700gr ring. Mix the liquid and powder then use a vacuum mixing machine to remove excess air from the mix. Invest the pattern and bench set for a minimum of 30 minutes, 20 minutes in and 10minutes out of the ring to dissipate moisture, the longer the better. However if putting the ring in the oven for overnight burnout place the rings in a plastic bag with a bit of water inside. Tie off the top with a burnable material. This will prevent drying of the ring and mitigate possible fracture of the mold.
Ensure adequate investment around each pattern. Typically 3-5mm between each pattern and 10mm from the sides of the ring and about 6mm from the top. This will allow for sufficient expansion in the mold and strength to minimize finning.
Once you have loaded your oven at room temperature set the rate of climb to 18F per minute to 800F and hold for 30 minutes. For the second stage, the rate of climb should be 25F per minute to 1650F (or the alloy manufacturer recommended temperature) and hold for 40 minutes. Add 10 minutes to final burnout time for each additional ring. Modifications to the liquid water ratio and final burnout temperature will vary for fixed work and pressable ceramics.
Can I replace the Power Switch on the Vibrators?
Unfortunately no. The power switch is not a ‘standalone’ switch but is attached directly to the electronic board in the machine.
I have an older Whip Mix Heavy Duty Vibrator with the slide lever in the front. It no longer makes the vibration get higher when I move the lever. Can I replace the part?
This model of Vibrator is discontinued and there are no longer parts available from Whip Mix.
You might try a hardware store or a local Dental Repair facility.
The New Heavy Duty Vibrator (ITEM# 10675) has several newer features that you will find beneficial. Please visit our Web Site at www.whipmix.com, click on equipment and scroll to VIBRATORS.
My older General Purpose Vibrator became very loud and then stopped working. What do I need to repair it?
If this is one of the original, Whip Mix General Purpose Vibrators this version has been discontinued. Unfortunately, parts are no longer available for this unit from Whip Mix. The new GP Vibrator (ITEM # 10650) is available through your dealer.
My Whip Mix Hygrobath isn’t working. Do you have spare parts?
Unfortunately, the Whip Mix Hygrobath (generally for use with Whip Mix Beauty Cast investment) is discontinued as of 2005 and no parts were available after September of 2012.
The Hygrobath held a water temperature of 38.7C or 100F
If you still wish to do the Hygroscopic Technique with your investment, we would suggest that you find a small water heater, candle heater or crock pot that you can control the temperature and hold at 100F.
The Vibrator is no longer working. What is wrong?
The Vibrator rod will wear out over time and needs replacing.
ITEM # 06424 Vibrator Rod
- Release the spring retainer under the vibrator arm and lift the arm.
- Grab the black Vibrator Rod with a hemostat or cotton plyers or if it is too short, some utility or Orthodontic wax.
- Pull the worn rod from the hole and replace with the new rod.
- Re-engage the spring retainer.
What is the ‘Vibration’ level of the Vibrators?
What is the best way to keep Stone off the Vibrator? It is hard to keep clean!
There have been many inventive Ways to keep the stone from getting on the vibrator from Paper towels to patient bibs held on with rubber bands.
The best way we have found is to cover the entire machine with a Hotel Shower Cap… These thin plastic caps with elastic retainers are in expensive (free from the hotel where you stay) and last a good while. Let the Stone drip and harden. At the end of the day remove the shower cap and shake. The stone will come right off. Replace the cap over the machine for the next days use.
What is the correct wiring configuration for the logic board for the Xcavator?
What is the power requirement or usage for the Vibrators?
Are there replacement Parts available for the VPMMini?
Below are the current (10/22/18) available replacement parts for the VPMMini.
ME5304 VPMMini Cordeset 115V- No item # Found
ME5305 VPMMini Cordeset 230V- Item# 29402
ME5306 Filter Black For VPMMini- Item# 29414
ME5307 VPMMini Silencer- Item# 29415
ME5308 Air Filter Element (Just Cartridge)- Item# 29416
ME5309 O-Ring VPMMini- Item# 29417
Checking the O ring on the vac-u-mixer did not fix my problem, how do I check the large black O ring on the edge of the Block?
Do you have a flexible mixing bowl?
How do I choose between the VPM2 or the VPMMini?
When choosing a hands free mixer that is programmable, our choices are the standard VPM2 or the VPM Mini. Both use the same Vac-U-Mixer assembly so this is a good aspect if the customer already has the VPM2 and does not want to have an inventory of obsolete accessories. Some of the benefits of going with the VPM2 is the ability to get more elaborate with the programming of the individual programs. Such as pre-vacuum, post vacuum, a premix and the ability to do staged mixing clockwise or counter clockwise with the option of varying RPM’s. The VPMMini will allow RPMs to be set from 100 to 500 and allow up to 9 minutes and 55 seconds of mix time in 5-second increments. The unit also stores up to 99 programs. Therefore, if a user is looking for the ability to create very specific programs, the VPM2 is a good choice. If however the user just wants a basic mix at a lower cost that is easy to program the VPMMini is the better option.
How do I lubricate the Shaft of the mixer bowl?
Most are unaware that this needs doing on occasion. Non-lubrication is often the cause of the tightness between the paddle and the lid.
- Remove the drive nut by grasping the paddle and then using a laboratory knife or a large screwdriver placed in the slot of the drive nut.
- Turn to loosen and then remove the drive nut.
- Now that the drive nut is off, pull the paddle shaft from the lid.
- You will notice two ‘O’ rings (#1 ‘O’ ring for the 200, 300, & 500 bowls) on the paddle shaft, you may wish to replace these.
- Then using Whip Mix Lubriplate (a white silicone grease) coat the paddle shaft and ‘O’ rings.
- Replace paddle shaft into the lid and replace the drive nut.
I get a dry spot in the middle of my mixing bowl. What causes this??
A low volume of mix material, in a higher volume bowl, causes this. In this case, it is less than 70grams of powder material in a 500ml bowl. The paddles are set just off the floor of the bowl. As they begin to spin, they push the material outward, leaving behind this dry spot. With more material in the mixing bowl, it folds in as it mixes and picks up the remainder of the dry material.
If you want to continue using this amount of material, order a 300ml mixing bowl system and it will solve the problem.
I have a Pro Blend mixer that needs parts.
Unfortunately, the PRO BLEND Mixer is discontinued as of 2006
All remaining parts were exhausted January 1, 2016.
There are no remaining parts.
I have an older Large Mixing Bowl that I use on my Mixer without the Vacuum ( #6800 Mechanical Mixer). It holds twice as much as the Stainless Steel bowl (6 inch Diameter Bowl and flexible). Where can I get replacements?
Unfortunately, this mixing bowl and the Mixing Lid with Paddle Shaft Assembly we discontinued many years ago. There are no replacements.
I have an Orthodontic Mixer with the Handy Holder. I need a new lid.
The lid for the Handy Holder is only available in the 500ml and the 875ml bowl types.
It is the same lid for both.
ITEM # 15482 – Plastic Lid with Locking Tabs and Collar for use with Handy Holder
I have hardened Stone in my mixing bowl. How do I get it out?
The best way is to soak the stone in the bowl with “GYP-STRIP’ a product from Whip Mix. This caustic material will soften and remove the stone without hurting the plastic of the bowl. It is also safe on the Plastic Lid and the Steel Mixing paddle and Shaft.
Caution: Gyp Strip should not be used on Aluminum parts or machines like Model Trimmers.
27014 Gyp-Strip 3.75 Liters (1 Gallon)
27022 Gyp-Strip 3.75 Liters (1 Gallon) Case of 4
I have sucked Stone up the tubing of my Mixing Machine. What do I do?
I just sucked water up into the machine and it is coming out the muffler! What do I do?
Unplug the machine….All is not lost.
- Remove the oil jar. Dispose of the oil and wipe out the Jar.
- Remove the Black Muffler from the machine and blow out the extra water with compressed air.
- Remove the silver cone from inside the oil jar and remove the white plastic wick holder by unscrewing.
- Remove the felt filters from off the white wick holder and blot dry on paper towel. Re-assemble the felts, wick holder and silver cone. Screw back into the oil jar lid.
- Remove the 6 Phillips head screws from the top of the machine and remove the top.
- Dry up all of the water in the pump area and then drop in a few drops of oil.
- Replace the top of the machine and replace the 6 screws.
- Replace the oil jar into the oil jar lid, with the new oil to the oil line.
- Plug in and turn on the machine. Run for one and one half minute, to lubricate and purge the machine of water vapor.
I need to replace the O ring on the Lid of the Vac-u-Mixer. What do I do?
I’m having trouble pulling vacuum. What do I do?
I’m not getting a good seal on my VPMMini and the bowl won’t hold, and/or the mix has bubbles.
Is the electrical Micro Leakage OK?
My bowl just fell and cracked. Can I just glue it?
Generally, you cannot. Super glue, which is thin enough to get into the open crack, will wash out over time. In addition, you may not get everything sealed up. The leak will allow air to enter into the mixing bowl and cause the loss of vacuum and possibly bubbles in your mix.
The bowls by themselves are a replaceable item for each size.
07595 – Plastic Bowl (200ml) 2-7/8 I.D. x 2” deep
07609 – Plastic Bowl (300ml) 2-7/8 I.D. x 3” deep
08974 – Plastic Bowl (500ml) 3-3/4 I. D. x 2-3/4 deep
09032 – Plastic Bowl (875ml) 3-3/4 I.D. x 5-1/4 deep
09067 – Stainless Steel (1200ml) 4-1/4 I.D. x 5-1/2 deep
My Lid and paddle just fell and it bent. Can I just bend it back?
In most cases the answer is no. Especially if the shaft has bent. These are very hard to straighten and if they are not straight, the paddles will strike the inside of the mixing bowl causing it to dislodge and loose vacuum and/or scrape bits of plastic into your mix.
A bent shaft also causes wear to the interior bearing of the lid. Eventually you will have an air leak or an actual seizure of the lid and the paddle shaft.
Replacement paddle shafts are available;
07617 – 200 ml Paddle Assembly (left hand thread for friction drive) with drive nut # 4109
08907 – 200 ml Paddle Assembly (right hand thread for slotted drive) with drive nut #6375
07633 – 300 ml Paddle Assembly (left hand thread for friction drive) with drive nut # 4109
08931 – 300 ml Paddle Assembly (right hand thread for slotted drive) with drive nut # 6375
08923 – 300 ml Paddle Assembly Teflon Coated for Alginates (right hand thread) drive nut # 6375
15601 – 500 ml Paddle Assembly (right hand thread, for use with handy holder) drive nut # 6393
08982 – 500 ml Paddle Assembly (right hand thread) drive nut # 6375
09008 – 500 ml Paddle Assembly (right hand thread, flat paddle for Alginate) drive nut # 6375
15644 – 875 ml Paddle Assembly (right hand thread, for use with Handy Holder) drive nut # 6393
09040 – 875 ml Paddle Assembly (right hand thread) drive nut # 6375
09075 – 1200 ml Paddle Assembly (right hand thread, for stainless steel bowl) drive nut # 6865
My machine came with 3 clean foam filters. Where do I put the 3rd?
My machine will not turn on. What is wrong?
My Powermixer Plus / Combination Unit is humming. What do I do?
From time to time, a Powermixer or Combination unit will seize up and make a humming sound. The good news is this is most likely not a motor issue but a rotor issue. The rotor can seize up due to moisture buildup in the pump area. One obvious reason for moisture in the pump is if water gets sucked up the hose and is allowed to sit. A less obvious reason is not purging the unit for one minute after each use. This is the most common reason and easiest to overlook. It is very important to get in the habit of purging after each use. Remember when we mix with vacuum in addition to pulling out air bubbles to get a bubble free mix, we are pulling moisture into the pump area. If we mix and simply stop the mixer the moisture has time to thicken in the oil causing vanes to stick, resulting in loss of vacuum and in some cases actual corrosion. When this happens, the rotor will cease to spin and the unit will not mix.
The sound from the motor will be a low hum giving the impression that the motor is ‘locked up’. Some users go so far as to try to turn the front drive chuck with a pair of pliers. Do not do this! While you may put enough force to free the rotor in the pump area, you may also strip the nylon gear in the lower housing.
First, unplug the unit from the wall outlet.
To access the pump area take the six screw off the top of the machine and remove the end plate. This will reveal the pump area. To do this you will need a Philips head screwdriver. Next you will want to remove the four vanes in the rotor and set them aside. Then take some WD40 and spray into just the pump area inside the pump body wall. Be careful not to spay down into the impellers in the top of the motor. Allow the WD40 to set for 15 minutes, longer if corrosion is extreme. Next take some fine grit sandpaper, 240-400 grit will work best, and fold it up to fit in the vane channels of the rotor. In addition to moving the sandpaper through the channel, hold the folded sandpaper down to make contact with the pump body floor. Once in contact with the floor try to turn the rotor via the front drive chuck at the bottom of the unit. REMINDER…unplug the machine. You can insert the drive nut of a lid and paddle assembly into the drive chuck, and use the paddle as a key to more easily turn the motor shaft. Once you feel movement, take out the paddle, plug the unit in and quickly turn on the switch to see if the drive chuck (s) spins. If they do spin, clean the vanes with WD40 and scrub them off.
Dip the vanes in oil (VV Lubricant) and insert them into the channels of the rotor making sure the notches of the vanes face the center of the rotor. Replace the end plate securing all six screws with even pressure to avoid a gap between the end plate and pump body wall, which can cause a loss in vacuum. If you are still experiencing low vacuum you may need to clean more thoroughly.
My Vacuum will not hold the Bowl. What is wrong?
If the 1200ml Vac-U-Mixer is not holding to the VPM2, you are not achieving full vacuum and the unit will not begin a mix. Checking for stone buildup in the foam filters and the ‘O’ring on the block assembly and the Vac-u-mixer lid could be the issue. The buildup of debris is most likely due to overfilling the mixing bowls. This causes the mixed material to come out of the vacuum hole on the lid and into the vacuum chamber. In addition, if the mixing bowl has been dropped, and landed on its lip, the lip may be bent and no longer capable of holding vacuum. This bowl should be replaced. In the 1200ml size the most common cause for vacuum issues is not holding the lid and bowl assembly square on the block at the lower housing of the VPM2. Also not holding it long enough to allow vacuum to be pulled, can be an issue. The greater volume of air needing removal from the 1200ml size will mean holding the bowl by hand longer compared to the 500ml for example. The vacuum pump only pulls at one rate and does not recognize the different sizes in use.
My VPM2 leaks. Should I be concerned?
Oil is blowing out of the black (silver in older versions) Muffler. What is wrong?
The bottom of my bowl has a number on it. What does that mean?
The drive is turning the paddles in the bowl but I have no vacuum.
The Machine just ‘humms’ when I turn it on and will not turn.
The “Vanes” in the top pump portion of the machine are stuck. You can see how to replace these vanes by going to the website, www.whipmix.com.
ITEM # 06467 Vanes for Vacuum Pump (set of 4)
You can unstick the vanes by following the first part of the video.
It may also be time to give this machine a maintenance checkup.
Purchase the Vac-U-Care Maintenance Kit, ITEM# 28398, and have all the parts necessary to get this unit back in shape.
The Mixing bowl lid: I dropped and cracked it… The connection to the tube is loose and I have replaced the small ‘O’ ring…. The paddle turns very hard and even squeaks when in use… I have an air leak…
What Oil can I use for these Machines?
ITEM# 06475 – VV Lubricant
The viscosity of the oil is important to the function of these machines. Do not use motor oil or 3 in 1 oil.
If you have an emergency and cannot wait to get the VV lubricant, call Technical Support (800-626-5651) for a substitute.
Why are there different Drive Nuts? Why don’t they all fit my paddle?
There have been different versions of the drive nut over the lifetimes and designs of the mixing machines based on the requirements of the Gypsum and Investment materials of the time.
Some of the older style drive nuts are no longer available.
The current Drive Nuts are these:
09105 – Slotted Drive Nut – for use with the 1200ml mixer (Stainless Steel). This drive nut fits the mixing machines but has a larger threaded area to accommodate the larger paddle shaft of the mixer.
09091 – Slotted Drive Nut for 875ml, 500ml, 300ml and 200ml mixing bowels
15555 – Short Slotted (hourglass shape) for use with the Handy Holder on the Ortho Power Mixer
09113 – Friction Grip (rubber ‘O’ ring around the center) for use with early mixing machines or the Combination Unit Mixer. This drive nut has left hand threads and can only be used for a paddle shaft that has left hand threads.
I just installed a new Electric Water Valve on my Model Trimmer, but now it is making a Thumping Noise….
A customer called with an issue regarding his newly installed 115V Electric Water Valve on a 1/2HP Model Trimmer. He said when the unit started up the water flowed just fine. However, when the unit is switched off, the water stopped but the water inlet hose vibrated and made a thumping noise. After consulting with LEAD Repair, they told me to have the customer check and make sure it was 115v and not 230v. They said this has caused the problem in the past. Also, double-check the tubing and make sure the water is coming into the ‘IN’ arrow on the brass base. The arrow indicates the direction of water flow. After re-checking the water tube, the customer realized the water inlet hose was going in the wrong direction. This was most likely causing his issue.
I want to replace the power switch on my ¾ hp trimmer. What is the part number?
What Gasket do I use for my Model Trimmer?
Care of Wax Pencils and Tips for the Digital Wax Carving Pencil
Some things to consider after you have familiarized yourself with the instructions of the Digital Wax Carving Pencil are to be aware of the unique features of the device to keep the unit in good working order.
When placing the tip on the pencil assembly, be sure to line up the slot in the tip base with the raised area of the heat element. Failure to do this and attempting to force the tip into place could result in damage to the tip and pencil. Additionally, while the pencil is not in use place it back in the magnetic cradle to keep it out of the way when performing other tasks. Leaving it resting on the bench could cause accidental burning of the electrical cord from the hot tip. Another problem occurs when the pencil falls on the floor, unknown to the user and the chair rolls over the handle breaking the assembly.
When changing tips it is easier to remove the tip while it is still warm. Using a pair of hemostats or needle nose pliers to prevent burning fingers, grasp the body of the tip and gently remove it from the heater rod. Once the wax residue cools, it may be very difficult to remove. Do not pull with excessive force as this can damage the heating element on the pencil.
Over time as tips are used, the insertion area may distort. This is normal. They can be gently bent back into position to provide a more snug fit to the heating element using a hemostat or needle nose plier. Excessive carbon and oxidation buildup on the tip and heating element created over time through heating and cooling cycles, is removed with glass beads in a micro blasting machine.
The tip will eventually wear out and need replacement. The tips are available for purchase individually.
Exercising a bit of care with the wax carving pencil can prolong the lifespan of the unit and avoid costly replacement or void the warranty.
Does this Digital Wax Pot Multi well work with both 220 and 115 Volt outlets?
Yes, you would just need to change the plug/electrical cord. The unit will adjust to the current.
How much wax will each of the wells hold in my Digital Wax Pot?
There are four (4) wells in the Digital Wax Pot. Opening the lid and looking from the front of the wax pot, the bottom well holds 62ml of wax. The two outside top wells hold 16ml of wax each, and the top center well holds 22ml of wax.
We cannot tell you how many grams of wax each well could contain, as each manufactured wax is a different weight.
I need to replace the Tip but it is stuck. What do I do?
Wax gets in between the tip and the heating rod of the electric waxer and when cool causes the wax tip to be ‘stuck’ to the rod.
Turn the waxer on and wait for it to heat. Using a pair of Hemostats, grab the tip. Do not rotate the tip as it is engaged in a lug and slot to lock it in place. Pull apart the tip and the pencil.
To replace with a new tip, find the lug on the heat rod and the slot in the new Tip. Line them up and then push together. The lug will engage the slot in the tip as you push them together and stop the tip from rotating in use.
I rolled over the pencil with my chair and broke it. Can I just replace the pencil?
Yes, you can! The pencils are available for the regular (ITEM# 10163) and for the TOUCH pencil (ITEM# 10165).
What do you mean by a “Touch Pencil”?
The touch feature allows you to “”Jump” between pre-set temperatures.
While holding the pencil, if you touch to metal ring at the top of the safety grip near the heated tip, it will cause the temperature of the pencil to jump to another pre-set temperature.
You may pre-set any two temperatures you would like, with-in the temperature range of the unit (50C-200C)
What is the best temperature setting for the Electric Waxer?
Each wax and each Technician will have a preference. Generally, for Crown & Bridge Wax, that temperature is around 195F (about 90C). The best way to tell if to get the tip hot enough to melt your individual wax is to gently press the heated tip into the wax. Make sure that it melts wax onto the tip. If not, it is too cool. Then rotate the tip of the waxer to a down pointing angle. The wax should move to the tip of the waxer and stay there.
IF it falls off, it is too hot….if it is smoking, it is too hot. If it does not melt when touching the tip to the wax, it is too cool.
What is this the Electric Heating Tool?
The Electric Heating Tool tool is used in Conjunction with a Surveyor for RPD (Removable Partial Denture) work.
This tool is discontinued by Whip Mix and is no longer available.
You may try META DENTAL or BesQual S542 or Henry Schein for availability.
Which Digital Wax Carving Pencil is the most popular?
Digital Wax Carving Pencils – These are what the Technicians simply refer to as Electric Waxers.
The Digital Wax Carving Pencil-Single is the most popular.
This unit has two temperature sets and can use all 9 available tips.
It has the smallest footprint (3×4.7×2.5”) and a Temperature range of 122F (50C) to 392F (200C)
Can I press e-Max with a Pro100 or a Pro200 Porcelain Furnace?
Yes. You may need to adjust the top temperature by 10C upward, and the press time should be about 3 minutes.
The Pro 200 has the IVOCLAR parameters already in the furnace. However, you will still need to make adjustments in the temperature. Remember that any porcelain reacts to time and temperature.
Can I press the new LiSi Press material from GC in a Pro Series press furnace?
Yes. The parameters provided to Whip Mix by GC are attached on page 2.
Remember that individual furnaces may require a slightly different number so be prepared to make slight changes to get your best results.
Do I have to use the Quick Cool Device on my PRO Porcelain Furnace?
No, you do not. The device is there to help cool the muffle in a more consistent fashion and faster than normal. Unfortunately, many Ceramists simply grab their airline and dispense compressed air into the Muffle to cool it rapidly. This is never a good idea as the compressed air may contain moisture, oil, and airline gunk, and you cannot actually regulate the amount of air entering the muffle. It will certainly shorten the life of the muffle and perhaps contaminate your ceramics.
The Quick cool device takes air from the muffle and routes it through the vacuum pump and back into the muffle safely and at a decreased temperature, cooling it more quickly than a natural cool.
If the battery dies on my PRO 100 Furnace what happens?
If the battery dies on your furnace while plugged into a power source nothing will happen.
However if the battery is dead and the power goes off, you will lose the use of the motherboard and be unable to restore your furnace. This is a permanent condition.
The PRO 100 uses 20+ year old (yes it has been that long since it was designed) technology for the retention of its files and instructions. Most of us are now using solid-state electronics that do not require battery backup and, we are not faced with this issue (the PRO 200 Furnaces are Solid State).
If you are using a PRO100 now and do not remember when you last changed the battery do not unplug your furnace! Call us for instructions about changing out the battery.
Why are my Zirconia Units coming out grey?
Recently a customer called concerned about sintered zirconia restorations coming out grey after sintering, using the ZR Pro discs. We went over firing parameters and all seemed normal in value. I recommended purging the unit at 50C over the highest sintering temperature, in this case 1500C. Once the purge process is complete, use calibration tabs to calibrate the sintering furnace. In some cases, temperatures can be off by 50C or more. This can affect the strength and visual esthetics of the restoration.
Additionally if you are using Alumina Sintering Beads, change them to Zirconia Sintering beads (ITEM # 96549).
I Have a Whip Mix Apollo burnout furnace. Can I still get parts?
Unfortunately, there are no parts available for this oven.
I have a Whip Mix Pro Burn 12 burn out oven. Can I still get parts?
Unfortunately no. This product was discontinued in 2005 and there were no parts or repairs as of September 2012.
I have an old Jelenko Burnout oven and the heater plate went out. I used to get these from Jelrus. Can I get heater plates from Whip Mix?
I heard Whip Mix bought Jelrus. Can you repair my Jelrus Porcelain Furnace?
Is it true that Whip Mix fixes the JELRUS burn out ovens?
Whip Mix acquired the Jelrus Burnout ovens from Air Techniques and we do repair or have parts available for some of the ovens. We have parts for the Jelrus Radiance, The Jelrus Infinity L-30 & M-30, and for the Jelrus PDQ Digital both Medium and Large. We do not repair or have parts available for the Jelrus Porcelain Furnaces.
Recently, I had a call from a customer looking for a Radiance Burn out furnace manual from Jelrus. Do we have these older furnace manuals?
I explained to the customer that we have limited literature and repair capability here at Whip Mix for the digital version of the Radiance Burnout Furnace from Jelrus.
However, the Jelrus machines we acquired from Air Techniques were the Infinity and P.D.Q. line and not the Radiance, which had been previously discontinued.
In addition, we did not acquire the Jelrus Porcelain Furnace lines, which were discontinued by Air Techniques.
Generally, we send customers to P&S in California as an alternate repair option, however I was able to locate a site that has many older or discontinued furnace manuals from several manufacturers available for download. The customer was very pleased to a have a source for these manuals.
Xcavator-Vacuum Fault Indication
From time to time when the Xcavator is running, you may see a red light illuminated in the vacuum fault area of the control panel. Even with this indicator light, your suction might still be operational.
Here are some things you will want to check if this occurs. The condition of your vacuum system, is every thing working properly? Then is the suction sufficient for use with the Xcavator? If multiple items are on the same vacuum line this can diminish the vacuum realized at the Xcavator connection. Sometimes closing off other items while using the Xcavator may be necessary.
Make sure the hose leading from the vacuum to the Xcavator is secure at both connection points. If suction seems weaker than usual, check the filter bag to see if it needs emptying. Sometimes the small black tube on the back of the unit will become obstructed. Removing the tube and manually blowing air through it may fix the issue. If you have cleaned the tube and reinstalled it yet still see the vacuum fault indicator light is illuminated, it may need to be reset.
To reset the vacuum fault indicator, you want to turn the vacuum and the Xcavator off. Next, hold down on the 200-gram delay button while turning on the power button for the Xcavator. You will hear a series of beeps, after which release the 200-gram delay button. Now turn off the Xcavator. You unit’s vacuum fault indicator should be reset.
Turn the Xcavator and the vacuum on. The Vacuum Fault light should now be out and the Xcavator should function normally. If it does not, you may need to redo the reset steps. If this still does not resolve the issue, please call Whip Mix technical support.